Repeatability and agreement between a reference Scheimpflug tomographer and a low-cost Scheimpflug system

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Eye Ctr, Beijing, Peoples R China [2]Royal Victorian Eye & Ear Hosp, Ctr Eye Res Australia, East Melbourne, Australia [3]Univ Melbourne, Dept Surg Ophthalmol, Melbourne, Vic, Australia
出处:
ISSN:

摘要:
Purpose: To assess the repeatability and agreement between Scheimpflug-based corneal topographers, Scansys and Pentacam, in measuring ocular parameters in myopic eyes. Setting: Grade-A tertiary hospital in Beijing, China. Design: Cross-sectional study. Methods: The following measurements were performed in 204 subjects undergoing preoperative examinations for refractive surgery from the corneal curvature, astigmatism, corneal thickness, corneal volume, corneal asphericity, anterior chamber depth ( ACD) and volume (ACV), and pupil diameters. Repeatability was determined using intraclass correlation coefficient with 95% CI, Cronbach a, coefficient of variation, within- subject SD, test-retest repeatability, and 1-way analysis of variance (1-way analysis of variance). The interdevice agreement was determined using paired t test and BlandAltman plots. Results: Scansys showed good repeatability in the anterior and posterior corneal flattest meridian (Kf), steepest meridian (Ks), and mean (Km); axis of anterior corneal Ks; anterior corneal astigmatism; pupil central corneal thickness (CCT); CCT; corneal apex thickness; thinnest corneal thickness; corneal volume; photopic pupil diameter; ACD; and ACV. In the agreement study, Bland-Altman plots showed that 95% limit of agreement of corneal curvature, pupil CCT, corneal apex thickness, thinnest corneal thickness, and corneal volume generated by Scansys and Pentacam were narrow. There were no statistically significant differences in the anterior corneal Kf and Km. Conclusions: Scansys showed good repeatability in measuring corneal curvature of anterior and posterior surfaces, anterior corneal astigmatism, corneal thickness, corneal volume, photopic pupil diameter, ACV, and ACD. Scansys and Pentacam can only be used interchangeably in measuring anterior corneal curvature. Copyright (c) 2023 Published by Wolters Kluwer on behalf of ASCRS and ESCRS

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2022]版:
大类 | 2 区 医学
小类 | 2 区 眼科学 2 区 外科
最新[2023]版:
大类 | 3 区 医学
小类 | 2 区 眼科学 3 区 外科
JCR分区:
出版当年[2021]版:
Q2 OPHTHALMOLOGY Q2 SURGERY
最新[2023]版:
Q1 SURGERY Q2 OPHTHALMOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2021版] 出版当年五年平均 出版前一年[2020版] 出版后一年[2022版]

第一作者:
第一作者机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Eye Ctr, Beijing, Peoples R China
通讯作者:
通讯机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Eye Ctr, Beijing, Peoples R China [*1]Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, No. 1Dong Jiao Min Xiang St, Dongcheng District, Beijing, 100730, China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21169 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)