Impacts of Adaptive Statistical Iterative Reconstruction-V and Deep Learning Image Reconstruction Algorithms on Robustness of CT Radiomics Features: Opportunity for Minimizing Radiomics Variability Among Scans of Different Dose Levels

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

机构: [1]Shanghai Jiao Tong Univ, Tongren Hosp, Dept Imaging, Sch Med, Shanghai 200336, Peoples R China [2]Shanghai Jiao Tong Univ, Ruijin Hosp, Dept Radiol, Sch Med, Shanghai 200025, Peoples R China [3]GE Healthcare, Computed Tomog Res Ctr, Beijing 100176, Peoples R China [4]GE Healthcare, Computed Tomog Res Ctr, Shanghai 201203, Peoples R China [5]Imperial Coll London, Dept Mat, South Kensington Campus, London SW7 2AZ, England [6]Haohua Technol Co Ltd, Shanghai 201100, Peoples R China
出处:
ISSN:

关键词: Deep learning Multidetector computed tomography Reproducibility of results Image enhancement Image reconstruction

摘要:
This study aims to investigate the influence of adaptive statistical iterative reconstruction-V (ASIR-V) and deep learning image reconstruction (DLIR) on CT radiomics feature robustness. A standardized phantom was scanned under single-energy CT (SECT) and dual-energy CT (DECT) modes at standard and low (20 and 10 mGy) dose levels. Images of SECT 120 kVp and corresponding DECT 120 kVp-like virtual monochromatic images were generated with filtered back-projection (FBP), ASIR-V at 40% (AV-40) and 100% (AV-100) blending levels, and DLIR algorithm at low (DLIR-L), medium (DLIR-M), and high (DLIR-H) strength levels. Ninety-four features were extracted via Pyradiomics. Reproducibility of features was calculated between standard and low dose levels, between reconstruction algorithms in reference to FBP images, and within scan mode, using intraclass correlation coefficient (ICC) and concordance correlation coefficient (CCC). The average percentage of features with ICC>0.90 and CCC>0.90 between the two dose levels was 21.28% and 20.75% in AV-40 images, and 39.90% and 35.11% in AV-100 images, respectively, and increased from 15.43 to 45.22% and from 15.43 to 44.15% with an increasing strength level of DLIR. The average percentage of features with ICC>0.90 and CCC>0.90 in reference to FBP images was 26.07% and 25.80% in AV-40 images, and 18.88% and 18.62% in AV-100 images, respectively, and decreased from 27.93 to 17.82% and from 27.66 to 17.29% with an increasing strength level of DLIR. DLIR and ASIR-V algorithms showed low reproducibility in reference to FBP images, while the high-strength DLIR algorithm provides an opportunity for minimizing radiomics variability due to dose reduction.

基金:
语种:
被引次数:
WOS:
PubmedID:
第一作者:
第一作者机构: [1]Shanghai Jiao Tong Univ, Tongren Hosp, Dept Imaging, Sch Med, Shanghai 200336, Peoples R China
共同第一作者:
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:25471 今日访问量:0 总访问量:1498 更新日期:2025-06-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)