Minimizing serum creatinine interference in automated laboratory analyses through the implementation of urea/creatinine ratio alerting mechanisms

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China [2]Department of General Surgery Gastrointestinal Group,Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China [3]Central Labortary ,Shanghai Xuhui Central Hosptial/Zhongshan-Xuhui Hospital, Fudan University, Shanghai 200031,China [4]Department of Laboratory Medicine, College of Health Science and Technology, Shanghai Jiao Tong University school of Medicine,Shanghai 200025,China
出处:
ISSN:

关键词: Serum creatinine Urea/creatinine ratio Drug interference Alkaline picrate assay Sarcosine oxidase enzymatic assay LC-IDMS/MS Total laboratory automation

摘要:
This study was designed to increase the precision and clinical significance of serum creatinine (Cr) measurements by refining the detection methodology through comprehensive laboratory automation and by employing the urea/creatinine ratio(UCR) to detect and mitigate drug interference.We measured the serum Cr concentration via the Siemens sarcosine oxidase enzymatic (SOE) assay. A substantial dataset comprising nearly 2 million records from 98,377 individuals was extracted to form the foundational database for serum UCR analysis. Using the Refine R package, we delineated the reference intervals for UCR. Our laboratory's automated system adeptly identified samples exhibiting aberrant UCR values, prompting a re-evaluation of Cr levels via the Siemens Jaffe (improved alkaline picrate) assay. These Cr measurements were subsequently corroborated against Beckman and Ortho SOE assays and benchmarked against liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS), the reference standard.The reference interval for the serum UCR ranged from 0.047 to 0.143(90% CI 0.046-0.048, 0.135-0.148).Among the 58 abnormal samples assessed, the mean deviation in Cr detection for the Siemens, Beckman, and Ortho SOE assays relative to the LC-IDMS/MS results were -61.05%,-53.97%, and -17.18%, respectively. The Siemens Jaffe (improved alkaline picrate) assay revealed a mean deviation of -3.10% compared with that of LC-IDMS/MS. These findings indicate that the improved alkaline picrate assay can accurately reflect the true Cr levels in serum samples affected by calcium dobesilate or etamsylate interference.The implementation of an automated UCR-based screening protocol has been shown to efficiently pinpoint samples with drug-related interference. This strategic approach not only decreases clinical risk but also guarantees the fidelity of serum Cr measurements without undermining the overall productivity of the laboratory's automated workflow.Copyright © 2025 Elsevier B.V. All rights reserved.

基金:
语种:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2025]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
最新[2025]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
JCR分区:
出版当年[2023]版:
Q2 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q2 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2023版] 出版当年五年平均 出版前一年[2022版]

第一作者:
第一作者机构: [1]Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China
共同第一作者:
通讯作者:
通讯机构: [1]Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China [4]Department of Laboratory Medicine, College of Health Science and Technology, Shanghai Jiao Tong University school of Medicine,Shanghai 200025,China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:25750 今日访问量:0 总访问量:1506 更新日期:2025-06-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)