机构:[1]Ophthalmic Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Sciences Key Lab, Beijing 100730, China首都医科大学附属北京同仁医院首都医科大学附属同仁医院
To compare the regularity and accuracy of laser in situ keratomileusis (LASIK) flaps created by the Ziemer FEMTO LDV "Classic" (Ziemer "Classic") and Ziemer FEMTO LDV Crystal Line femtosecond laser (Ziemer Crystal Line).Fourier-domain optical coherence tomography (RTVue OCT) was used to measure the morphology of 200 LASIK flaps of 100 consecutive patients created with the Ziemer Classic (100 flaps) or the Ziemer Crystal Line (100 flaps) at one week postoperatively. Flap thickness was evaluated at 36 specified measurement points on each flap. For all procedures with both lasers, the nominal flap thickness was 110µm.The mean flap thickness of the Ziemer Crystal Line group (102.49±2.68µm) was thinner than that of the Ziemer Classic group (107.65±5.09µm) (P<0.01). Average thickness of all flaps was uniform within 4µm at all measurement points. The flaps in the Ziemer Crystal Line group were more regular than those in the Ziemer Classic group when measured from the center to the periphery. The maximum deviation from the nominal 110µm of 36 measurements was 8µm in the Ziemer Classic group, while in the Ziemer Crystal Line group it was 9µm. Within the 3 600 measurements on the 100 eyes, differences greater than 20µm were observed 0.14% in the Ziemer Classic group, and 0.04% in the Ziemer Crystal Line group.The flaps created with the Ziemer FEMTO LDV Crystal Line femtosecond laser are more uniform and thinner than those created by the Ziemer FEMTO LDV Classic femtosecond laser.
第一作者机构:[1]Ophthalmic Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Sciences Key Lab, Beijing 100730, China
通讯作者:
推荐引用方式(GB/T 7714):
Jing Zhang,Yue-Hua Zhou,Lei Tian,et al.Comparison of Ziemer FEMTO LDV Classic and Crystal Line femtosecond laser flap quality by Fourier-domain optical coherence tomography[J].INTERNATIONAL JOURNAL OF OPHTHALMOLOGY.2013,6(5):611-617.doi:10.3980/j.issn.2222-3959.2013.05.11.
APA:
Jing Zhang,Yue-Hua Zhou,Lei Tian&Chang-Bin Zhai.(2013).Comparison of Ziemer FEMTO LDV Classic and Crystal Line femtosecond laser flap quality by Fourier-domain optical coherence tomography.INTERNATIONAL JOURNAL OF OPHTHALMOLOGY,6,(5)
MLA:
Jing Zhang,et al."Comparison of Ziemer FEMTO LDV Classic and Crystal Line femtosecond laser flap quality by Fourier-domain optical coherence tomography".INTERNATIONAL JOURNAL OF OPHTHALMOLOGY 6..5(2013):611-617