高级检索
当前位置: 首页 > 详情页

Radiation Dose to the Lens Using Different Temporal Bone CT Scanning Protocols

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Capital Med Univ, Dept Radiol, Beijing Tongren Hosp, Beijing 100730, Peoples R China [2]Philips Healthcare, Beijing, Peoples R China
出处:
ISSN:

摘要:
BACKGROUND AND PURPOSE: Temporal bone CT is performed frequently in clinical practice. At the same time, the eye lens are exposed to ionizing radiation without any useful diagnostic information delivered. Our aim was to investigate the radiation dose to the lens of the eye by using temporal bone CT scanning with different protocols. MATERIALS AND METHODS: Direct axial and coronal CT by using a conventional sequential scanning mode (140 kV, 220 mAs/section, 1.25-mm thickness, 1.25-mm increment, 16 x 0.625 mm collimation, the glabellomeatal line as a scanning baseline), a routine helical scanning mode (140 kV, 220 mAs/section, 0.315 pitch, 0.67-mm thickness, 0.33-mm increment, 16 X 0.625 collimation, the orbitomeatal line as a scanning baseline), and a modified helical scanning mode (acanthiomeatal line as a scanning baseline; other parameters, same as above) was performed on an exsomatized cadaveric head. CTDl(vol) and DLP were recorded for each scanning mode, and effective doses were calculated. Organ doses for the lens were measured with TLDs. RESULTS: When the sequential scanning mode was used, the gross effective dose was 1.21 mSv and the organ dose to the lens was 50.96 and 1.73 mGy, respectively, for direct axial and coronal imaging. The effective dose was 0.803 mSv in routine helical scanning, while the lens dose was 40.17 mGy. With the modified helical scanning mode, the effective dose was as same as that for the routine helical scanning, but the lens dose was reduced significantly to 10.33 mGy. CONCLUSIONS: The effective doses resulting from sequential axial and coronal scanning were 1.51 times higher than the dose from helical scanning, and the lens dose was 1.31 times higher. With the modified helical scanning mode, thinner section images could be used to reformat axial, coronal, and sagittal images with a further 74.3% reduction in lens dose beyond that achieved with the conventional helical protocol.

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2009]版:
大类 | 3 区 医学
小类 | 3 区 临床神经病学 3 区 神经成像 3 区 核医学
最新[2025]版:
大类 | 2 区 医学
小类 | 2 区 核医学 3 区 临床神经病学 3 区 神经成像
JCR分区:
出版当年[2008]版:
Q1 NEUROIMAGING Q2 CLINICAL NEUROLOGY Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
最新[2023]版:
Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Q2 CLINICAL NEUROLOGY Q2 NEUROIMAGING

影响因子: 最新[2023版] 最新五年平均 出版当年[2008版] 出版当年五年平均 出版前一年[2007版] 出版后一年[2009版]

第一作者:
第一作者机构: [1]Capital Med Univ, Dept Radiol, Beijing Tongren Hosp, Beijing 100730, Peoples R China
通讯作者:
通讯机构: [1]Capital Med Univ, Dept Radiol, Beijing Tongren Hosp, Beijing 100730, Peoples R China [*1]Capital Med Univ, Dept Radiol, Beijing Tongren Hosp, 1 Dongjiaominxiang, Beijing 100730, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:25463 今日访问量:0 总访问量:1498 更新日期:2025-06-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)