Comparisons between Neural Response Imaging thresholds, electrically evoked auditory reflex thresholds and most comfortable loudness levels in CII Bionic Ear users with HiResolution (TM) sound processing strategies

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Capital Univ Med Sci, Beijing Tongren Hosp, Beijing Inst Otorhinolaryngol, Beijing 100730, Peoples R China
出处:
ISSN:

关键词: cochlear implants programming

摘要:
Conclusions. The data collected in this study indicated that first Neural Response Imaging (NRI) thresholds had a better correlation with HiResolution(TM) most comfortable loudness (M) levels than tNRI thresholds. Electrically evoked auditory reflex thresholds (EARTs) had a higher correlation with HiResolution M levels than tNRI thresholds and a lower correlation than first NRI thresholds. NRI is a very useful method for programming the cochlear implants of young children who cannot demonstrate a reliable judgment of loudness. Objective. To investigate how HiResolution sound processing, designed to deliver high-rate stimuli, relates to EARTs and electrically evoked compound action potential measurements produced by low-rate stimuli. Material and methods. Nine profoundly hearing-impaired children and adults aged 6-29 years participated in the study. NRI responses were elicited using pulse trains consisting of biphasic pulses at a pulse width per phase of 32 mu s delivered at a frequency of 30 Hz using SoundWave programming software. Stimuli were delivered to the odd electrodes (1, 3, 5, 7, 9, 11, 13 and 15) along the array. tNRI (NRI threshold) and first NRI thresholds were recorded for each stimulating electrode. "Speech bursts'' stimuli used in EARTs recording were delivered to four electrodes at a time and stapedial reflexes were recorded from the impedance bridge. The M levels used were those used by each patient in their everyday HiResolution programs. Results. For 8 patients (53 stimulating electrodes) the correlation between tNRI threshold and M level was r = 0.675 (p = 0.000) and that between first NRI thresholds and M level was r = 0.741 (p = 0.000). On average the M-level value was 20 CU (Current Unit) lower than the first NRI threshold value and 12 CU higher than the tNRI threshold value. The M-level patterns across the electrode array overall were similar to the tNRI or first NRI threshold patterns. For 7 patients (112 stimulating electrodes) the correlation between EART and M levels was r = 0.710 (p = 0.000). On average the EART value was 14 CU higher than the M-level value.

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2004]版:
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 耳鼻喉科学
JCR分区:
出版当年[2003]版:
Q2 OTORHINOLARYNGOLOGY
最新[2023]版:
Q3 OTORHINOLARYNGOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2003版] 出版当年五年平均 出版前一年[2002版] 出版后一年[2004版]

第一作者:
第一作者机构: [1]Capital Univ Med Sci, Beijing Tongren Hosp, Beijing Inst Otorhinolaryngol, Beijing 100730, Peoples R China
通讯作者:
通讯机构: [1]Capital Univ Med Sci, Beijing Tongren Hosp, Beijing Inst Otorhinolaryngol, Beijing 100730, Peoples R China [*1]Capital Univ Med Sci, Beijing Tongren Hosp, Beijing Inst Otorhinolaryngol, 17 Hougou Lane,Chongnei St, Beijing 100730, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:23549 今日访问量:0 总访问量:1282 更新日期:2025-04-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)