Comparability of thyroid-stimulating hormone immunoassays using fresh frozen human sera and external quality assessment data.

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Department of Clinical Laboratory, Beijing Chaoyang Hospital, Beijing Center for Clinical Laboratories, The Third Clinical Medical College of Capital Medical University, Beijing, P.R. China, [2]Department of Clinical Laboratory, Wangjing Hospital, China Academy of Traditional Chinese Medicine, Beijing, P.R. China, [3]Department of Clinical Laboratory, China-Japan Friendship Hospital, Beijing, P.R. China, [4]Department of Clinical Laboratory, Beijing Tongren Hospital, Capital Medical University, Beijing, P.R. China, [5]National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Beijing, P.R. China, [6]Department of Clinical Laboratory, Beijing Hospital, Beijing, P.R. China, [7]Department of Clinical Laboratory, Beijing Chaoyang Hospital, The Third Clinical Medical College of Capital Medical University, Beijing, P.R. China
出处:
ISSN:

摘要:
This study aimed to assess the comparability among assays using freshly frozen human sera and external quality assessment (EQA) data in China.Twenty-nine serum samples and two commercial EQA materials, obtained from the National Center for Clinical Laboratories (NCCL), were analyzed in triplicate using eight routine TSH assays. The commutability of commercial EQA materials (NCCL materials) was evaluated in accordance with the CLSI EP30-A and IFCC bias analysis. Median values obtained for the NCCL EQA materials were used to determine the systematic and commutability-related biases among immunoassays through back-calculation. The comparability of TSH measurements from a panel of clinical samples and NCCL EQA data was determined on the basis of Passing-Bablok regression. Furthermore, human serum pools were used to perform commutable EQA.NCCL EQA materials displayed commutability among three or five of seven assay combinations according CLSI or IFCC approach, respectively. The mean of systematic bias ranged from -13.78% to 9.85% for the eight routine TSH assays. After correcting for systematic bias, averaged commutability-related biases ranged between -42.26% and 12.19%. After correction for systematic and commutability -related biases, the slopes indicating interassay relatedness ranged from 0.801 to 1.299 using individual human sera, from 0.735 to 1.254 using NCCL EQA data, and from 0.729 to 1.115 using pooled human serum EQA(the commutable EQA).The harmonization of TSH measurement is challenging; hence, systematic and commutability-related biases should be determined and corrected for accurate comparisons among assays when using human individual serum and the commercial EQA materials.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2020]版:
大类 | 3 区 综合性期刊
小类 | 3 区 综合性期刊
最新[2023]版:
大类 | 3 区 综合性期刊
小类 | 3 区 综合性期刊
JCR分区:
出版当年[2019]版:
Q2 MULTIDISCIPLINARY SCIENCES
最新[2023]版:
Q1 MULTIDISCIPLINARY SCIENCES

影响因子: 最新[2023版] 最新五年平均 出版当年[2019版] 出版当年五年平均 出版前一年[2018版] 出版后一年[2020版]

第一作者:
第一作者机构: [1]Department of Clinical Laboratory, Beijing Chaoyang Hospital, Beijing Center for Clinical Laboratories, The Third Clinical Medical College of Capital Medical University, Beijing, P.R. China,
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21173 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)