Monte Carlo calculation of conversion coefficients for dose estimation in mammography based on a 3D detailed breast model

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China [2]Tsinghua Univ, Key Lab Particle & Radiat Imaging, Minist Educ, Beijing 100084, Peoples R China [3]Nuctech Co Ltd, Beijing 100084, Peoples R China [4]Captial Med Univ, Beijing Tongren Hosp, Beijing 100730, Peoples R China
出处:
ISSN:

关键词: conversion coefficient detailed breast model mammography mean glandular dose Monte Carlo

摘要:
Purpose: At present, the Chinese specification for testing of quality control in x-ray mammography is based on a simple breast model, and does not consider the glandular tissue distribution in the breast. In order to more precisely estimate the mean glandular dose (MGD) in mammography for Chinese women, a three-dimensional (3D) detailed breast model based on realistic structures in the breast and Chinese female breast parameters was built and applied in this study. Methods: To characterize the Chinese female breast, Chinese female breast parameters including breast size, compressed breast thickness (CBT), and glandular content were investigated in this study. A mathematical model with the detailed breast structures was constructed based on the Chinese female breast parameters. The mathematical model was then converted to a voxel model with voxels. The voxel model was compressed in craniocaudal (CC) view to obtain a deformation model. The compressed breast model was combined with the Chinese reference adult female whole-body voxel phantom (CRAF) to study the effects of backscatter from the female body. Monte Carlo simulations of the glandular dose in mammography were performed with Geant 4. The glandular tissue dose conversion coefficients for breasts with different glandular contents (5%, 25%, 50%, 75%, and 100% glandularity) and CBTs (3 cm, 4 cm, 5 cm, and 6 cm) were calculated, respectively, at various x-ray tube voltages (25 kV, 28 kV, 30 kV, 32 kV, and 35 kV) for various target/filter combinations (Mo/ Mo, Mo/Rh, Rh/Rh, and W/Rh). Results: A series of glandular tissue dose conversion coefficients for dose estimation in mammography were calculated. The conversion coefficients calculated in this study were compared with those estimated with the simple breast model. A discrepancy of 5.4-38.0% was observed. This was consistent with the results obtained from the realistic breast models in the literature. Conclusions: A 3D detailed breast model with realistic structures in the breast was constructed based on the Chinese female breast parameters. It was used to calculate glandular tissue dose conversion coefficients for mammography. Although the accuracy of the proposed model could not be directly assessed, the consistency of the obtained results with previously performed analyses increases the confidence in the applicability of the proposed model. The proposed model could be used in dose estimation and dose optimization for mammography of Chinese women. (C) 2017 American Association of Physicists in Medicine

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2016]版:
大类 | 3 区 医学
小类 | 3 区 核医学
最新[2023]版:
大类 | 2 区 医学
小类 | 3 区 核医学
JCR分区:
出版当年[2015]版:
Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
最新[2023]版:
Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING

影响因子: 最新[2023版] 最新五年平均 出版当年[2015版] 出版当年五年平均 出版前一年[2014版] 出版后一年[2016版]

第一作者:
第一作者机构: [1]Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China [2]Tsinghua Univ, Key Lab Particle & Radiat Imaging, Minist Educ, Beijing 100084, Peoples R China
通讯作者:
通讯机构: [1]Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China [2]Tsinghua Univ, Key Lab Particle & Radiat Imaging, Minist Educ, Beijing 100084, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21169 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)