Purpose: To investigate the settings for the optimal microwave ablation geometry with the simultaneous application of double 915 MHz antennae in ex vivo bovine livers, so as to provide the technical basis for treating large liver tumor in one ablation session. Materials and methods: MWAs were performed on ex vivo bovine livers by simultaneously application of double 915 MHz internally cooled-shaft antennae. Four power settings (50, 60, 70 and 80W) were used during MWAs, while application time was fixed at 10 min. Three inter-antenna distances (2.0, 2.5 and 3.0 cm) were used. Diameters and shapes of the coagulation zones were observed on gross specimens. Results: (1) The coagulation shape was related to the inter-antenna distance, which was most spherical at an inter-antenna distance of 2.0 cm. A recess of the coagulation zone was observed at an inter-antenna distances of 2.5 and 3.0 cm. (2) The long-axis and short-axis coagulation diameter enlarged with increasing power output. However, there were no significant differences in the coagulation diameters between 70 and 80W (P > 0.05). More desirable coagulation geometry could be obtained by simultaneous application of double antennae at 70 W for 10 min with an inter-antenna distance of 2.0 cm, the long-axis and short-axis coagulation diameter were 6.95 +/- 0.32 cm and 5.30 +/- 0.22 cm, respectively. Conclusion: Simultaneous application of double 915 MHz antennae can generate large coagulation zones with desirable shape which may be advantageous for treating large liver tumor in one ablation session. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
基金:
Natural Science Foundation of ChinaNational Natural Science Foundation of China (NSFC) [30825010]
第一作者机构:[1]Chinese Peoples Liberat Army Gen Hosp, Dept Intervent Ultrasound, Beijing 100853, Peoples R China[2]Capital Med Univ, Beijing TongRen Hosp, Dept Ultrasound, Beijing 100730, Peoples R China
通讯作者:
通讯机构:[1]Chinese Peoples Liberat Army Gen Hosp, Dept Intervent Ultrasound, Beijing 100853, Peoples R China[*1]Chinese Peoples Liberat Army Gen Hosp, Dept Intervent Ultrasound, 28 Fuxing Rd, Beijing 100853, Peoples R China
推荐引用方式(GB/T 7714):
Shi Wenyuan,Liang Ping,Zhu Qiang,et al.Microwave ablation: Results with double 915 MHz antennae in ex vivo bovine Livers[J].EUROPEAN JOURNAL OF RADIOLOGY.2011,79(2):214-217.doi:10.1016/j.ejrad.2010.03.015.
APA:
Shi, Wenyuan,Liang, Ping,Zhu, Qiang,Yu, Xiaoling,Shao, Qiujie...&Dong, Baowei.(2011).Microwave ablation: Results with double 915 MHz antennae in ex vivo bovine Livers.EUROPEAN JOURNAL OF RADIOLOGY,79,(2)
MLA:
Shi, Wenyuan,et al."Microwave ablation: Results with double 915 MHz antennae in ex vivo bovine Livers".EUROPEAN JOURNAL OF RADIOLOGY 79..2(2011):214-217