Objective: To evaluate the long-term safety and efficacy of the Clip coupler attached to the stapes head in patients with unilateral congenital aural atresia (CAA). Methods: This single-center retrospective study included 16 Mandarin-speaking patients who had unilateral microtia accompanied by CAA. All patients were divided into two groups: the short-term follow-up group (n = 9) and the long-term follow-up group (n = 7). The floating mass transducer of the Vibrant Soundbridge (VSB) was positioned in the stapes head by the Clip coupler. The safety of the VSB was investigated by comparing preoperative and postoperative bone-conduction (BC) thresholds as well as by complications. The effectiveness was evaluated by functional gain (FG), word recognition score (WRS), speech reception threshold (SRT) and signal-to-noise ratio (SNR). Results: Pre- and post-operative BC thresholds were no different in all patients. And no complications developed. VSB-aided thresholds in the free-field had improved significantly in both short- and long-term follow-up groups. The improvements of WRS were observed in two groups. The monosyllabic VSB-aided WRS in the long-term follow-up group was significantly higher than that in the short-term follow-up group. When speech was from the impaired ear and noise presented to the side of normal ear (SVSBNCL), lower SNRs were found in two groups after VSB implantation. However, there was no statistical difference in aided SNR between the two groups at SVSBNCL status. Conclusions: Our results show that the FMT connected to the stapes head is a secure and useful device for patients with unilateral CHL/MHL, not only in terms of improved hearing thresholds, but also improved speech intelligibility in quiet and noisy environments.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2019]版:
大类 | 4 区 医学
小类 | 3 区 耳鼻喉科学
最新[2023]版:
大类 | 4 区 医学
小类 | 4 区 耳鼻喉科学
JCR分区:
出版当年[2018]版:
Q3 OTORHINOLARYNGOLOGY
最新[2023]版:
Q3 OTORHINOLARYNGOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2018版] 出版当年五年平均 出版前一年[2017版] 出版后一年[2019版]

第一作者:
第一作者机构: [1]Capital Med Univ, Beijing Tongren Hosp, Dept Otolaryngol Head & Neck Surg, 1 Dong Jiao Min Xiang, Beijing 100730, Peoples R China
通讯作者:
通讯机构: [1]Capital Med Univ, Beijing Tongren Hosp, Dept Otolaryngol Head & Neck Surg, 1 Dong Jiao Min Xiang, Beijing 100730, Peoples R China [*1]Department of Otolaryngology Head and Neck Surgery, Beijing Tongren Hospital, Capital Medical University, No.1 Dong Jiao Min Xiang, Dongcheng District, Beijing 100730, China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21710 今日访问量:0 总访问量:1245 更新日期:2025-03-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)