Diagnostic performance of the automated breast volume scanner: a systematic review of inter-rater reliability/agreement and meta-analysis of diagnostic accuracy for differentiating benign and malignant breast lesions
机构:[1]Department of Ultrasound in Medicine, Shanghai Jiao Tong University Affiliated Sixth People’s Hospital, Shanghai Institute of Ultrasound in Medicine, No. 600 Yishan Road, Shanghai 200233, China[2]Department of Gerontology, Shanghai Jiao Tong University Affiliated Sixth People’s Hospital, No. 600 Yishan Road, Shanghai 200233, China[3]Department of Ultrasound in Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, No. 1111 Xianxia Road, Shanghai 200336, China
To investigate the inter-rater reliability and agreement of the automated breast volume scanner (ABVS) and the diagnostic accuracy for differentiating malignant and benign lesions. The overall aim was to find out if the ABVS is applicable to daily clinical practice. Qualifying studies were retrieved from Pubmed, EMBASE, Cochrane Library, Biosis Preview, CBM disc and by manual search and reference lists up to 30 September 2014. Pooled sensitivity and specificity of ABVS were calculated and summary receiver operating characteristic curves were drawn. Thirteen studies were included in the meta-analysis of diagnostic accuracy and seven studies were included in the systematic review of inter-rater reliability/agreement of ABVS. For 'diagnostic accuracy', the pooled values of sensitivity, specificity, positive and negative likelihood ratios, and diagnostic odds ratio were 92 % (95 % CI 89.9-93.8), 84.9 % (82.4-87.0), 6.172 (4.364-8.730), 0.101 (0.075-0.136), and 72.226 (39.637-131.61), respectively. For the studies of inter-rater reliability/agreement, the quality was heterogeneous and no evidenced result can be pooled. Sensitivity and specificity of ABVS for differentiating malignant and benign breast lesions were high. More sound studies focusing on inter-rater reliability/agreement of ABVS, which deeply affect the clinical utilization and generalization of ABVS, are urgently needed. aEuro cent ABVS has high sensitivity and specificity in differentiating malignant and benign breast lesions. aEuro cent The quality of published inter-rater reliability studies is heterogeneous. aEuro cent Empirical evidence concerning the inter-rater reliability/agreement for the ABVS is rare. aEuro cent Comparison studies on ABVS and other medical imaging examinations are warranted.
第一作者机构:[1]Department of Ultrasound in Medicine, Shanghai Jiao Tong University Affiliated Sixth People’s Hospital, Shanghai Institute of Ultrasound in Medicine, No. 600 Yishan Road, Shanghai 200233, China
共同第一作者:
通讯作者:
推荐引用方式(GB/T 7714):
Meng Zheying,Chen Cui,Zhu Yitong,et al.Diagnostic performance of the automated breast volume scanner: a systematic review of inter-rater reliability/agreement and meta-analysis of diagnostic accuracy for differentiating benign and malignant breast lesions[J].EUROPEAN RADIOLOGY.2015,25(12):3638-3647.doi:10.1007/s00330-015-3759-3.
APA:
Meng, Zheying,Chen, Cui,Zhu, Yitong,Zhang, Shuling,Wei, Cong...&Shen, E..(2015).Diagnostic performance of the automated breast volume scanner: a systematic review of inter-rater reliability/agreement and meta-analysis of diagnostic accuracy for differentiating benign and malignant breast lesions.EUROPEAN RADIOLOGY,25,(12)
MLA:
Meng, Zheying,et al."Diagnostic performance of the automated breast volume scanner: a systematic review of inter-rater reliability/agreement and meta-analysis of diagnostic accuracy for differentiating benign and malignant breast lesions".EUROPEAN RADIOLOGY 25..12(2015):3638-3647