A light efficiency uniformity detection system for Medical Rigid Endoscope based on image processing

文献详情

资源类型:
WOS体系:

收录情况: ◇ CPCI(ISTP) ◇ EI

机构: [1]Beijing Inst Technol, Sch Optoelect, Beijing Key Lab Precis Optoelect Measurement Inst, Beijing 100081, Peoples R China [2]State Food & Drug Adm, Beijing Ctr Med Device Qual Supervis & Testing, Beijing Inst Med Device Testing, Beijing 101111, Peoples R China [3]Capital Med Univ, Beijing Tongren Hosp, Beijing 100730, Peoples R China
出处:
ISSN:

关键词: medical rigid endoscope light efficiency uniformity image processing plot curve center axisymmetric field of view (FOV) Synthetically relative effect of edge light energy

摘要:
Light efficiency uniformity is a very important parameter of medical rigid endoscope. This paper introduces a new system based on image processing to test the light efficiency uniformity of medical rigid endoscope. Employing an electric machinery to reduce the human intervention, so that the precision of measuring and automation degree are improved. We collect the image with a digital CCD camera and display it on the screen of a computer, which can avoid visual fatigue from the direct observation through the rigid endoscope. To perform the image processing on a computer, we adopt a self-developed image processing software, by which the test results can be obtained from PC itself. The processes of our self-developed image processing software include: gray-scale transformation, image pretreatment and image binarization; calculate the center and equivalent radius of the field of view (FOV); plot the curve, through which the ratio of edge and center in different field and the center axisymmetric of light efficiency can be both calculated. It concludes that the relative self-effect of illumination light luminosity is the foremost factor affecting the uniformity, and these endoscopes are all qualified with the max deviation of the center axisymmetric less than 20%. The results of our study prove that this system can test the light efficiency uniformity of medical rigid endoscope quickly, expediently and accurately, and it contains more information instead of only reflecting a particular field of the FOV, what's more, it applies to different types, length and angles of view of medical rigid endoscope.

基金:
语种:
WOS:
第一作者:
第一作者机构: [1]Beijing Inst Technol, Sch Optoelect, Beijing Key Lab Precis Optoelect Measurement Inst, Beijing 100081, Peoples R China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21173 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)