Reproducibility of Anterior Chamber Angle Measurement Using the Tongren Ultrasound Biomicroscopy Analysis System

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Beijing Ophthalmology & Visual Science Key Lab, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing [2]Department of Ophthalmology, Liaocheng Clinical School, Taishan Medical College, Liaocheng People’s Hospital, Taishan Medical University, Taian [3]Department of Ophthalmology, Handan 3rd Hospital, Handan, People’s Republic of China [4]Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, Hong Kong [5]Queensland Eye Institute & University of Queensland, Queensland, Australia.
出处:
ISSN:

关键词: reproducibility anterior chamber angle ultrasound biomcroscopy

摘要:
Purpose:To assess the intraobserver and interobserver variability of new software for the analysis of anterior chamber-angle images obtained using ultrasound biomicroscopy (UBM).Patients and Methods:Consecutive UBM images of 4 quadrants (the superior, the nasal, the temporal, and the inferior) of 30 primary angle-closure suspects (PACSs) and 30 normal subjects with open angles were selected. Two observers used the new UBM image analysis software to calculate the following parameters on 2 separate occasions: angle opening distance (AOD), trabecular-iris angle (TIA), angle recess area (ARA), trabecular-ciliary process distance (TCPD), iris-ciliary process distance (ICPD), and iris thickness (IT). Intraobserver and interobserver reproducibility were measured using intraclass correlation coefficient (ICC) and Bland-Altman plots.Results:For intraobserver measurements in PACS, the range of ICC for AODs, TIAs, ARAs, TCPDs, ICPDs, and ITs in the 4 quadrants were 0.77 to 0.94, 0.80 to 0.92, 0.82 to 0.94, 0.51 to 0.73, 0.42 to 0.78, and 0.76 to 0.89; in open angles, these were 0.83 to 0.94, 0.80 to 0.93, 0.84 to 0.93, 0.57 to 0.89, 0.57 to 0.84, and 0.70 to 0.95. For interobserver measurements in PACS, the range of ICC for AODs, TIAs, ARAs, TCPDs, ICPDs, and ITs in the same quadrants as above were 0.52 to 0.89, 0.65 to 0.92, 0.68 to 0.92, 0.22 to 0.80, 0.48 to 0.93, and 0.29 to 0.75; in open angles, these were 0.57 to 0.87, 0.60 to 0.86, 0.64 to 0.90, 0.27 to 0.94, 0.56 to 0.93, and 0.28 to 0.66.Conclusions:Angle width (AOD, TIA, and ARA) and IT of both PACS and open angles could be reliably measured by the same observer; interobserver measurements and those related to the ciliary process (TCPD and ICPD) were more variable. For the current version of this software, we recommend measurements of anterior chamber parameters by the same observer.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2013]版:
大类 | 4 区 医学
小类 | 3 区 眼科学
最新[2023]版:
大类 | 4 区 医学
小类 | 4 区 眼科学
JCR分区:
出版当年[2012]版:
Q2 OPHTHALMOLOGY
最新[2023]版:
Q2 OPHTHALMOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2012版] 出版当年五年平均 出版前一年[2011版] 出版后一年[2013版]

第一作者:
第一作者机构: [1]Beijing Ophthalmology & Visual Science Key Lab, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing
共同第一作者:
通讯作者:
通讯机构: [1]Beijing Ophthalmology & Visual Science Key Lab, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing [*1]Beijing Ophthalmology & Visual Science Key Lab, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, No. 1 Dong Jiao Min Xiang Street, Dongcheng District, Beijing 100730, People’s Republic of China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21169 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)