Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Inst Ophthalmol,Beijing Tongren Eye Ctr, Beijing 100730, Peoples R China
出处:
ISSN:

关键词: repeatability agreement Visante AS-OCT Casia SS-1000 OCT implantable collamer

摘要:
AIM: To evaluate the intra-operator repeatability of time domain and swept-source Fourier domain anterior segment optical coherence tomography (AS-OCT), namely, Visante AS-OCT and Casia 55-1000 OCT, in measuring the preoperative parameters of implantable collamer lens (ICL) in myopic eyes, as well as the agreement between the two devices. METHODS: A total of 97 eyes from 49 myopes were investigated in this prospective case series study. The anterior chamber depth (ACD), angle-to-angle distance (ATA), pupil diameter (PD) and crystalline lens rise (CLR) in all subjects were measured for three times during one session by the same operator. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability limits and intraclass correlation coefficients (ICC). The agreement between the two systems was evaluated using the Bland-Altman plots and 95% limits of agreement (LoA). RESULTS: The repeatability limits of Visante AS-OCT in measuring ACD, ATA, PD and CLR were 0.099, 0.141, 0.304, and 0.079 mm, respectively. The repeatability limits of Casia SS-1000 OCT in measuring ACD, ATA, PD, and CLR were 0.105, 0.127, 0.357, and 0.082 mm, respectively. Excellent repeatability could be attained in both devices, with the ICC>0.8 for all the measured variables. The interdevice agreement was excellent (P>0.05) for ACD and ATA, but poor (P<0.05) for PD and CLR. CONCLUSION: Good repeatability can be attained by time domain and swept-source Fourier-domain OCT for all the measured variables. Moreover, interdevice agreement analysis suggests that interchangeable measurements between two devices can be achieved for ACD and ATA, but not for PD and CLR; but the differences in measurements were not clinically significant.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2019]版:
大类 | 3 区 医学
小类 | 3 区 眼科学
最新[2023]版:
大类 | 4 区 医学
小类 | 4 区 眼科学
JCR分区:
出版当年[2018]版:
Q4 OPHTHALMOLOGY
最新[2023]版:
Q2 OPHTHALMOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2018版] 出版当年五年平均 出版前一年[2017版] 出版后一年[2019版]

第一作者:
第一作者机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Inst Ophthalmol,Beijing Tongren Eye Ctr, Beijing 100730, Peoples R China
通讯作者:
通讯机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Inst Ophthalmol,Beijing Tongren Eye Ctr, Beijing 100730, Peoples R China [*1]Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Institute of Ophthalmology, Beijing Ophthalmology and Visual Sciences Key Laboratory, Beijing 100730, China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21169 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)