Reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Capital Med Univ, Beijing Ophthalmol & Vis Sci Key Lab, Beijing Tongren Eye Ctr, Beijing Inst Ophthalmol,Beijing Tongren Hosp, Beijing 100730, Peoples R China
出处:
ISSN:

关键词: optical quality double-pass system best focus repeatability reproducibility

摘要:
AIM: To evaluate intro -session repeatability and reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system. METHODS: Thirty Chinese healthy adults (19 to 40 years old) meeting our inclusion criterion were enrolled in the study. After a basic eye examination, two methods of optical quality measurement, based on subjective and objective best focuses were performed using the Optical Quality Analysis System (OQAS) with an artificial pupil diameter of 4.0 mm. RESULTS: With each method, three consecutive measurements of the following parameters: the modulation transfer function cutoff frequency (MTFcutoff), the Strehl(2D) ratio, the OQAS values (OVs) at contrasts of 100%, 20%, 9% and the objective scatter index (OSI) were performed by an experienced examiner. The repeatability of each method was evaluated by the repeatability limit (AL) and the coefficient of repeatability (COR). Reproducibility of the two methods was evaluated by intra -class correlation coefficient (ICC) and the 95% limits of agreement (Bland and Altman analysis). Thirty subjects, seven females and twenty three males, of whom 15 right eyes and 15 left eyes were selected randomly for recruitment in the study. The RLs (percentage) for the six parameters measured at objective focus and subjective focus ranged from 8.44% to 15.13% and 10.85% to 16.26%, respectively. The CORs for the two measurement methods ranged from 8.27% to 14.83% and 10.63% to 15.93%, respectively. With regard to reproducibility, the ICCs for the six parameters of OQAS ranged from 0.024 to 0.276. The 95% limits of agreement obtained for the six parameters (in comparison of the two methods) ranged from -0.57 to 42.18 (MTFcutoff), -0.01 to 0.23 (Strehl(2D) ratio), -0.02 to 1.40 (OV100%), -0.10 to 1.75 (OV20%), -0.14 to 1.80 (OV9%) and -1.46 to 0.18 (OSI). CONCLUSION: Measurements provided by OQAS with either method showed a good repeatability. However, the results obtained from the two different measurement methods showed a poor reproducibility. These findings suggest that it might be best to evaluate patients' optical quality by OCIAS using the best focus as chosen automatically by the instrument.

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2014]版:
大类 | 4 区 医学
小类 | 4 区 眼科学
最新[2023]版:
大类 | 4 区 医学
小类 | 4 区 眼科学
JCR分区:
出版当年[2013]版:
Q4 OPHTHALMOLOGY
最新[2023]版:
Q2 OPHTHALMOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2013版] 出版当年五年平均 出版前一年[2012版] 出版后一年[2014版]

第一作者:
第一作者机构: [1]Capital Med Univ, Beijing Ophthalmol & Vis Sci Key Lab, Beijing Tongren Eye Ctr, Beijing Inst Ophthalmol,Beijing Tongren Hosp, Beijing 100730, Peoples R China [*1]Capital Med Univ, Beijing Ophthalmol & Vis Sci Key Lab, Beijing Tongren Eye Ctr, Beijing Inst Ophthalmol,Beijing Tongren Hosp, 1 Dong Jiao Min Xiang, Beijing 100730, Peoples R China
通讯作者:
通讯机构: [1]Capital Med Univ, Beijing Ophthalmol & Vis Sci Key Lab, Beijing Tongren Eye Ctr, Beijing Inst Ophthalmol,Beijing Tongren Hosp, Beijing 100730, Peoples R China [*1]Capital Med Univ, Beijing Ophthalmol & Vis Sci Key Lab, Beijing Tongren Eye Ctr, Beijing Inst Ophthalmol,Beijing Tongren Hosp, 1 Dong Jiao Min Xiang, Beijing 100730, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21166 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)