Background Copernicus optical coherence tomography (SOOT) is a new, ultra high-speed and high-resolution instrument available for clinical evaluation of optic nerve. The purpose of the study was to compare the agreements between SOOT and Heidelberg retinal tomography (HRT). Methods A total of 44 healthy normal volunteers were recruited in this study. One eye in each subject was selected randomly. Agreement between SOOT and HRT-3 in measuring optic disc area was assessed using Bland-Altman plots. Relationships between measurements of optic nerve head parameter obtained by SOOT and HRT-3 were assessed by Pearson correlation. Results There was no significant difference in the average cup area (0.306 vs. 0.355 mm, P=0.766), cup volume (0.158 vs. 0.130 mm, P=0.106) and cup/disc ration (0.394 vs. 0.349 mm, P=0.576) measured by the two instruments. However, other optic disc parameters from SOOT were significantly lower compared with HRT-3. The Bland-Altman plot revealed good agreement of cup area and cup volume measured by SOOT and HRT-3. Bad agreement of disc area, rim area, rim volume and cup/disc ratio were found between SOOT and HRT-3. The highest correlations between the two instruments were observed for cup area (r(2)=0.783, P=0.000) and cup/disc ratio (r(2)=0.669, P=0.000), whereas the lowest correlation was observed for disc area (r(2)=0.100, P=0.037), rim area (r(2)=0.275, P=0.000), cup volume (r(2)=0.005, P=0.391) and rim volume (r(2)=0.021, P=0.346). Conclusions There were poor agreements between SOOT and HRT-3 for measurement of optic nerve parameters except cup area and cup volume. Measurement results of the two instruments are not interchangeable. Chin Med J 2012;125(16):2858-2861
第一作者机构:[1]Capital Med Univ, Beijing Tongren Hosp, Beijing Tongren Eye Ctr, Beijing Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China[2]Nanjing Tongren Hosp, Dept Ophthalmol, Nanjing 211102, Jiangsu, Peoples R China
通讯作者:
通讯机构:[1]Capital Med Univ, Beijing Tongren Hosp, Beijing Tongren Eye Ctr, Beijing Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China[2]Nanjing Tongren Hosp, Dept Ophthalmol, Nanjing 211102, Jiangsu, Peoples R China
推荐引用方式(GB/T 7714):
Yang Qing-song,Yu Ya-jie,Li Shu-ning,et al.Comparative study of optic disc measurement by Copernicus optical coherence tomography and Heidelberg retinal tomography[J].CHINESE MEDICAL JOURNAL.2012,125(16):2858-2861.doi:10.3760/cma.j.issn.0366-6999.2012.16.011.
APA:
Yang Qing-song,Yu Ya-jie,Li Shu-ning,Liu Juan&Hao Ying-juan.(2012).Comparative study of optic disc measurement by Copernicus optical coherence tomography and Heidelberg retinal tomography.CHINESE MEDICAL JOURNAL,125,(16)
MLA:
Yang Qing-song,et al."Comparative study of optic disc measurement by Copernicus optical coherence tomography and Heidelberg retinal tomography".CHINESE MEDICAL JOURNAL 125..16(2012):2858-2861