Comparative study of optic disc measurement by Copernicus optical coherence tomography and Heidelberg retinal tomography

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE ◇ 统计源期刊 ◇ CSCD-C ◇ 中华系列

机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Tongren Eye Ctr, Beijing Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China [2]Nanjing Tongren Hosp, Dept Ophthalmol, Nanjing 211102, Jiangsu, Peoples R China
出处:
ISSN:

关键词: Copernicus optical coherence tomography Heidelberg retinal tomography-3 agreement

摘要:
Background Copernicus optical coherence tomography (SOOT) is a new, ultra high-speed and high-resolution instrument available for clinical evaluation of optic nerve. The purpose of the study was to compare the agreements between SOOT and Heidelberg retinal tomography (HRT). Methods A total of 44 healthy normal volunteers were recruited in this study. One eye in each subject was selected randomly. Agreement between SOOT and HRT-3 in measuring optic disc area was assessed using Bland-Altman plots. Relationships between measurements of optic nerve head parameter obtained by SOOT and HRT-3 were assessed by Pearson correlation. Results There was no significant difference in the average cup area (0.306 vs. 0.355 mm, P=0.766), cup volume (0.158 vs. 0.130 mm, P=0.106) and cup/disc ration (0.394 vs. 0.349 mm, P=0.576) measured by the two instruments. However, other optic disc parameters from SOOT were significantly lower compared with HRT-3. The Bland-Altman plot revealed good agreement of cup area and cup volume measured by SOOT and HRT-3. Bad agreement of disc area, rim area, rim volume and cup/disc ratio were found between SOOT and HRT-3. The highest correlations between the two instruments were observed for cup area (r(2)=0.783, P=0.000) and cup/disc ratio (r(2)=0.669, P=0.000), whereas the lowest correlation was observed for disc area (r(2)=0.100, P=0.037), rim area (r(2)=0.275, P=0.000), cup volume (r(2)=0.005, P=0.391) and rim volume (r(2)=0.021, P=0.346). Conclusions There were poor agreements between SOOT and HRT-3 for measurement of optic nerve parameters except cup area and cup volume. Measurement results of the two instruments are not interchangeable. Chin Med J 2012;125(16):2858-2861

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2011]版:
大类 | 4 区 医学
小类 | 4 区 医学:内科
最新[2023]版:
大类 | 3 区 医学
小类 | 3 区 医学:内科
JCR分区:
出版当年[2010]版:
Q3 MEDICINE, GENERAL & INTERNAL
最新[2023]版:
Q1 MEDICINE, GENERAL & INTERNAL

影响因子: 最新[2023版] 最新五年平均 出版当年[2010版] 出版当年五年平均 出版前一年[2009版] 出版后一年[2011版]

第一作者:
第一作者机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Tongren Eye Ctr, Beijing Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China [2]Nanjing Tongren Hosp, Dept Ophthalmol, Nanjing 211102, Jiangsu, Peoples R China
通讯作者:
通讯机构: [1]Capital Med Univ, Beijing Tongren Hosp, Beijing Tongren Eye Ctr, Beijing Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China [2]Nanjing Tongren Hosp, Dept Ophthalmol, Nanjing 211102, Jiangsu, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:21169 今日访问量:0 总访问量:1219 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学附属北京同仁医院 技术支持:重庆聚合科技有限公司 地址:北京市东城区东交民巷1号(100730)